Logo

Manufacturer/company:

Automated Precision Europe GmbH

Im Breitspiel 17
Heidelberg
Germany

Number of projects
1
Number of test pieces
1
Phone

06221 7298050

Email

https://apimetrology.com/contact-api/

Website

https://apimetrology.com/

Documents

AP 2601-1710_statement_01.pdf
Valid until 13.02.2031
AP 2601-1710_certificate_01.pdf
Valid until 13.02.2031
AP 2601-1710_Urkunde_01.pdf
Valid until 13.02.2031
AP 2601-1710_Bescheinigung_01.pdf
Valid until 13.02.2031

Test piece

Name of test piece/s API Radian PRO
Description Laser power: 1 mW; wavelength: IFM 632.8 nm
Project number AP 2601-1710
Created on 02/20/2026
Valid until 02/13/2031
Test type Particle emission (ISO 14644-1, -14)
Classfication Suitability for air cleanliness class: (ISO-Class):1
Particle emission (ISO 14644-1, -14)

Particle emission (ISO 14644-1, -14)

Test parameter(s)

Air Cleanliness Class

Cycle time: 19.121 s
Break time: 0 s
Horizontal travel distance: 148.6321
Vertical travel distance: 173.1588°
Rotational speed horizontal: 0.741 rad/s
Rotational speed vertical: 0.861 rad/s

1

Overall result

1


Please note: Transport damages, incorrect installation, oil leakage, aging behavior, corrosion etc. can influence the test result.