Test piece

Name of test piece/s API Radian PRO
Description Laser power: 1 mW; wavelength: IFM 632.8 nm
Project number AP 2601-1710
Created on 02/20/2026
Valid until 02/13/2031
Test type Riboflavin test Equipments (VDMA leaflet “Riboflavin test for low-germ or sterile process technologies - Fluorescence test for testing cleanability”)
Classfication weak
Riboflavin test Equipments (VDMA leaflet “Riboflavin test for low-germ or sterile process technologies - Fluorescence test for testing cleanability”)
The examination of cleanability of API Radian PRO was investigated according to VDMA information sheet. The following test result could could be provided:

Classification

weak


The API Radian PRO can only be cleaned to a limited extent by using a simple wiping procedure with ultra-pure water. The fluorescence test identified several critical areas. It is extremely difficult to clean these areas effectively. These areas have to be cleaned especially thoroughly or using a more complex procedure, e.g. by removing certain parts before cleaning. The ability to clean the API Radian PRO effectively could be improved by eliminating most of the depressions or undercuts